top of page

Specifications

NEUROS CEREBRAL OXIMETRY

StO2 Measurement Range:

Trending Accuracy of StO2:

Resolution of StO2:

Display Update Rate:

 

VENUS 2000 CVP

CVP Measurement Range:

Accuracy:

Resolution:

Sensitivity:

Display Update Rate:

 

VO 100 JUGULAR VENOUS OXIMETRY

SjvO2 Measurement Range:

Accuracy of SjvO2:

Resolution of StO2:

Display Update Rate:

 

ELECTRICAL

Power Draw:

Voltage Input:

Measurement in NIR Wavelength:

NIR Power:

 

I/O Signals:

 

ENVIRONMENTAL

Temperature:

 

Humidity:

 

Air Pressure:

 

 

COMPLIANCE

Safety Standard for Medical Equipment:

EMC Standard:

Degree of Protection:

Mode of Operation:

​

DB 9-PIN CONNECTOR

NEUROS CEREBRAL OXIMETRY

A0 - 100%

A±1.5%

A1%

A1 second

​

VENUS 2000 CVP

A0 - 20 mmHg or 0 - 27 cmH2O

A±2.16 mmHg or ±2.94 cmH2O
A±1 mmHg or ±1 cmH2O

A±1 mmHg or ±1 cmH2O

A1 second

​

VO 100 JUGULAR VENOUS OXIMETRY

A0 - 100%

A± 2.7%

A1%

A1 second

​

ELECTRICAL

A200 mW typical operation @ 5.0VDC input voltage

A+5.0VDC ± 0.25VDC, w/50mV max. ripple

A905±5nm, 740±4nm, and 660±3nm

APeak: 14 mW

AAverage: 0.9 mW

A0 to +5.0VDC (nominal) @ +5.0VDC input voltage

​

ENVIRONMENTAL

AOperating: 10 - 38 degrees Celsius

ATransportation/Storage: -20 - 50 degrees Celsius

AOperating: 30 - 70 %

ATransportation/Storage: 20 - 80%

AOperating: 86 - 106 kPa

ATransportation/Storage: 86 - 106 kPa

A

COMPLIANCE

AIEC 60601-1 edition 3 Amendment 1

AIEC 60601-1-2 Class A

AType BF-Applied Part

AContinuous

DB9 PIN Description.png